Fully automatic sterilization from start to the end with high pressure steam sterilization followed by “Drying” process
Features of Steam Sterilizer (Laboratory Use, with Drying Process) (SM201 *Limited product for overseas)
Fully automatic sterilization from start to the end with high pressure steam sterilization followed by “Drying” process
Features of Steam Sterilizer (Laboratory Use, with Drying Process) (SM201 *Limited product for overseas)
Feature
Automatic operation from sterilization to drying process can be easily set by the interactive key input system.
Features
Quick sample drying function makes samples ready to use right after sterilization.
Features
Drying temperature can be set depending on the application within the range from 150 to 180℃.FeaturesThe water level in drain bottle can be easily checked with the level indicator on the front panel.
Model | SM201 *Limited product for overseas |
---|---|
System | Automatic high-pressure steam sterilization |
Operating temp. range (Sterilization) | 105~123℃ |
Operating temp. range (Drying) | 150~180℃ |
Max. pressure capacity | 0.18MPa |
Safety device | Self-diagnostic functions(Detection sensor error, SSR short circuit, Heater disconnection, Main relay error), Safety valve, ELB, Drain bottle switch |
Accessories | Stainless steel mesh basket (205dia. x 204Dmm) x 2 pcs., Drain board, Chemical indicator strip, Drain bottle, Condensation collection container with magnetic bracket, each 1 pc. |
Interior | Stainless steel |
Temp. Controller | PID control by microprocessor |
Temp. Display | Digital display by green LED and setting by UP/DOWN key |
Timer/Resolution | 1 min. ~99hrs. and 59min./100~999hrs./1min. or 1hr. |
Internal dimensions(φ×H mm) | φ240×455mm |
External dimensions(W×D×H mm) | 410×470×957mm |
Remarks(dimensions) | 20L |
Power Source | 115V 13A 00 |
Weight | 65kg |