Dry bath (incubator) HB-R48 1 block 5°C to 95°C

Suitable for:

  • Enzyme reactions, serum inactivation, incubation and other laboratory procedures

Description

Technical features:

  • Adjustment function for adjusting the displayed temperature to the actual temperature (±10.0 °C)
  • 2 Peltier cooling modules
  • Unit manufactured from stainless steel and powder coated steel, anodized aluminum block
  • including removal tongs “BLE101” and block cover
  • Heater (300W), PT100 sensor
  • Save temperature and timer settings
  • Backlit LCD display
  • CE certificate and traceability through serial numbers

 

Safety device:

  • lockdown mode
  • Overheat and overcurrent protection
  • Sensor error detection

 

Steering:

  • digital PID and fuzzy control with jog shuttle switch

 

Unit: Dry bath HB-R48 unit, power cable, user manual
Set: Dry bath HB-R48 unit, block BLC548 for 48x 1.5/2 ml reaction tubes, forceps BLE101, power cable, user manual

features

Devices with a mains voltage of 230 V can also be supplied with a voltage of 120 V on request.

width (inside)

90mm

length (inside)

120mm

height (inside)

60mm

width (outside)

185mm

length (outside)

230mm

height (outside)

220mm

cooling time

2.3°C/min. between 95°C and 25°C, 0.5°C/min. between 25°C and -5°C

heat-up time

approx. 5°C/min

power consumption

450W

temperature range

Room temperature -5°C to 95°C

temperature accuracy

±0.1 at 37°C, ±0.2°C at 90°C

packaging size

350x220x310mm